Since coming back from Germany for HARRIS I've been chasing a weird issue on my ...
Andrew Zonenberg @azonenberg@ioc.exchange
Since coming back from Germany for HARRIS I've been chasing a weird issue on my ThunderScope 100baseT1 test setup.
Before leaving for HARRIS I unplugged the TS, left the rest of the setup untouched, and brought the TS with me to demo to people while traveling.
After coming back the TS appears fine (passed calibration, and gives good-looking eyes on 100baseTX).
But when hooked back to the same, theoretically unmodified baseT1 test setup, I get weird diagonal distortion, lots of what I think is mostly common mode noise that bleeds a bit into differential mode too, and eye closure.
Before and after. Same ngscopeclient version, same session file, same DUT, same scope. What changed?