Adventures in troubleshooting. Everyone needs a good clock.
So the device has a 10MHz clock with a 1ppm TXCO and a 3.3V powered buffered output driver rated to -40C. According to the NIST traceable calibrated frequency counter, the 10MHz is 8.4Vp-p and off by a few kilohertz, sometimes jumping up to 20MHz below -40C. So I hook up the BNC connector to the digital oscilloscope and what do we have, a nice low noise square wave at exactly 10.000MHz at 3.03Vp-p, quite a different story than the frequency counter. But maybe the scope is missing some erratic noise not shown. But the trigger would have picked that up, especially at 8.4Vpp the frequency counter claimed. A different scope backs up the first scope. Two calibrated scopes versus a calibrated frequency counter. Since I have several of these calibrated frequency counters and scopes, they all share the same conflict. Who is right, the operator or the equipment?